AR-5500B Two million digital gastrointestinal DRF-DR X-ray System, Quality Medical Equipment from China-ARI Medical Equipment of High Quality - ARI Medical Technology Co., Ltd. - ARI Group

Products

AR-5500B

Two million digital gastrointestinal DRF


Digital Radiography and Digital Fluoroscopy System

I.  Application

Full support perspective, gastrointestinal spot film, GI (barium meal, barium enema), orthopedic photography, pediatrics photography, chest, urinary system angiography, peripheral angiography operations, gynecological Photography (HSG) and many other checks, the real machine can realize all these function.

II.  Standard configure

1

X ray Tube

E7252X

1Unti

Toshiba

2

Electrical Collimator

XSQ20

1Unit

3

Generator

FSQ60

1Unit

4

Flat Panel Detector

1417V

1Unit

5

Diagnostic table

ZDC20

1Unit

6

Console

KZT40

1Unit

7

Intensifier

E5830SD-P6A

1Unit

Toshiba

8

Image                 acquisition

workstation

DSI60

1Unit

9

Dedicated camera

acA1920-40gm

1Set

BASLER


III.  Specification

Item

Content

Technical Parameters

Power Supply

Voltage

380V±38V

Frequency

50Hz±1Hz

Capacity

≥65kVA

Internal Resistance

≤0.17Ω

Generator (FSQ60)

Power

56KW

Inverter Frequency

440 KHz

Radiography Tube voltage

40kV150kV step adjustment

Radiography Tube current

10mA710mA step adjustment

Radiography Time

1.0ms10000ms step adjustment

Radiography mAs

0.1 mAs900 mAs

Fluoroscopy Tube voltage

40kV125kV step adjustment

Fluoroscopy Tube current

0.5mA10mA continuous fluoroscopy

5mA20mA Pulse fluoroscopy

X-ray tube assembly

E7252X

Target

Molybdenum-based lanthanum-tungsten

composite

Target angle

12°

Nominal tube voltage

150kV

Tube Focus: big/small

1.2mm / 0.6mm

Input power

Big Focus/ 75kW     Small focus/ 27kW

Anode thermal capacity

210kJ(300KHU)

Anode maximum heat

dissipation

475W667KHU/min

Component heat capacity

900kJ1250KHU

Anode Rotation Speed

9700rpm

Collimator field of view light

Halogen lamp, AC24V/100W

Collimator

Visible light illumination

Average illumination brightness: >100Lux

Light field exposure time

5-45s, 5s per step

Diagnostic

Table transverse movement

220mm

table diagnostic table

distance

SID

1100mm1500mm

Table panel equivalent

filtration

≤1.2mmAl

Point device movement range

720mm

Table height from the ground

900mm

Table size

2100mm×796mm

Table rotation

90°~ -25°

Minimum distance of the compressor from the table surface

≤150mm

Compressor pressure

80N130N

Table bearing

135kg

Fragmentation

Whole film, two-piece, four-slice

Fixed grid

Grid density: 103L/INCH, Ratio: 10:1,

convergence distance 130 cm stationary type :15” ×18”

Field of view size

230mm

Intensifier (E5830SD- P6A)

Resolution

52 Lp/cm

Conversion factor

cd / m2

 SHAPE  \* MERGEFORMAT

26 uGy / s

X-ray camera (acA1920-

40gm)

Signal to noise ratio

45 dB

Cell size

5.86×5.86μm

Number of valid pixels

1920×1200

Wi-Fi Flat Panel Detector (1417V)

Image Sensor

Amorphous silicon thin film transistor

Pixel size

150μm

Effective pixel size

2304H*2800V

Effective area(H x V)

343H*416V

Image transmission

IEEE802.11a/b/g/n

Gray scale

14bit

Spatial resolution

3.4 Lp/mmCsI

Energy range

40kV -150kVP

Power input

DC24V

1, computer: B150 PLUS motherboard;

Image

processor: Intel Core I5 6400

acquisition

Memory: 8 GB DDR3

workstation

Computer system

Hard disk: 1T SATA (7200 rpm)

(DSI60)

Optical drive: DVR-221CHV CD burner

NIC: Intel(R)gigabit CT

Graphics card: gtx1050

Monitor

Dell (DELL) U2412Mc display

Workstation software

Basic functions: new patient; new check; anonymous registration; WorkList search and advanced search; WorkList refresh; WorkList information import; check list; view image; film print;

The collection part is divided into gastrointestinal collection and DR collection functions: Gastrointestinal collection function: bed selection; patient shape selection; specific examination protocol selection; gastrointestinal and contrast protocol selection; selected protocol list; real-time image noise reduction selection; real-time image enhancement selection; single frame storage; continuous perspective storage; Frame exposure storage; sequence exposure storage; real-time image horizontal flipping; real-time image vertical flipping; generator perspective state switching; perspective time accumulation; fluoroscopy time clearing; ABS function switch;

DR acquisition function: window width window position adjustment; 90° rotation to the left; 90° rotation to the right; horizontal mirroring; vertical mirroring; image negative; cropping frame; exposure parameter setting; stadium focus switching; exposure mode selection; ionization

chamber selection ; exposure actual dose value; system status indicator.

Image viewing function: image zooming; partial zooming; original size; recovery window size; image grid display; 90° left rotation; 90° rotation to the right; horizontal mirroring; vertical mirroring; text logo; line logo; angle logo; Marking; cropping frame; rectangle marking; circle marking; polygon marking; length correction; four-corner information display; ruler display; window width window level adjustment; sequence image playback; image output; silhouette function; deleting image; image enhancement; ; image negative.

Film printing function: add image; insert image; delete image; clear image; image layout; image selection; print format; film print setting; film format setting; add, modify, delete printer

settings.

IV. Work station

Integrated combined operation handle, can easily control the bed body, image system and the movement of rotary pedal, spot film;

1) The table body can +90 degrees to 0 to -25 degrees rotate.

2) Humanized design of diagnosis bedside switch operation, can control the table body and imaging system movement,so that the close table inspection is convenient and easy operation;

3) Spot film device and imaging system movement range more than 720 mm,

4) Adopts the operation of the machine move, but the patient don't need move ,, can easily finish from the throat, esophagus to the lower abdomen of a series of inspections; 5)Cassette trolley can test cassette size by itself, a key can completely finish the insert piece, save space, convenient and fast;



Copyright © 2001-2020 ARI Medical Technology Co., Ltd. - ARI Technology Group Co., Ltd. All Rights Reserved.