AR-Y200

Infant Ventilator

Introduction

-AR-Y200 ventilator is a new-type of pneumatic-driven and electronic-controlled ventilator which is designed to meet the special needs of clinical respiratory management and respiratory therapy for infants and children with respiratory disorders. With the feature of innovative functions, convenient operation, less gas consumption, compact but reasonable structure, and outstanding safety and reliability, it is an indispensable instrument in emergency and ICU rooms, and an essential tool to rescue the life of infants and children.

-AR-Y200 ventilator can work with some innovative ventilation modes, e.g., constant flow

mode, pressure limit mode and time cycle ventilation mode, and also supports PEEP and IMV functions.

Technical Specifications

PARAMETER

SPECIFICATION

Working Mode (Ventilation Mode)

Controlled Mechanical Ventilation (CMV),

Intermittent Mandatory Ventilation (IMV),

Continuous Positive Airway Pressure (CPAP),

Manual Ventilation (Manual).

Breath Rate

1~120BPM

Inspiratory to Expiratory Ratio

1:1~1:3

Special Order I:E Range

10:1~1:9.9

Inspiratory Flow

3 L/min~20 L/min

Tidal Volume

Minimum 10 mL~300 mL

Minute Volume

Max. Ventilation volume greater than 5 L/min

Oxygen Concentration of Output Gas

21%~100%

Ventilator System Compliance

Not greater than 3×10-2 mL/Pa

Positive End-Expiratory Pressure (PEEP)

0kPa~2.0kPa

Inhalation Safety Valve

Safety valve shall be opened when the airway pressure is within -300 Pa ~ -800 Pa

Inspiratory Pressure

1kPa~9kPa

Inspiratory Time

0.2s~3.0s

Synchronous Trigger

Pressure Trigger

-1.0~0kPa (Special Order)

Flow Trigger

0~10L (Special Order)

Maximum Safe Pressure

Not greater than 10kPa

Continuous Positive Airway Pressure (CPAP)

0kPa~2.0kPa

Monitoring Tidal Volume

Real-time monitoring within the adjustment range of tidal volume: 100 mL: ±10 mL; 100 mL: ±15 %;

Monitoring Oxygen Concentration

21%~100%

Power Failure Alarm

Audible alarm sounds immediately in case of power outage and lasts at least for 120s

Gas Source Failure Alarm

Audible and visual alarm sounds within 6s after gas source interruption occurs and lasts at least for 120s

Airway Pressure Lower Limit Alarm Range

PEEP +0.2kPa

Airway Pressure Upper Limit Alarm Range

Inspiratory pressure +0.4 kPa